ECG beat detection using filter banks.
Afonso VX, Tompkins WJ, Nguyen TQ, Luo S.
Endocardial Solutions, Inc., Saint Paul, MN 55108, USA.
We have designed a multirate digital signal processing algorithm to detect heart beats in the electrocardiogram (ECG). The algorithm incorporates a filter bank (FB) which decomposes the ECG into subbands with uniform frequency bandwidths. The FB-based algorithm enables independent time and frequency analysis to be performed on a signal. Features computed from a set of the subbands and a heuristic detection strategy are used to fuse decisions from multiple one-channel beat detection algorithms. The overall beat detection algorithm has a sensitivity of 99.59% and a positive predictivity of 99.56% against the MIT/BIH database. Furthermore this is a real-time algorithm since its beat detection latency is minimal. The FB-based beat detection algorithm also inherently lends itself to a computationally efficient structure since the detection logic operates at the subband rate. The FB-based structure is potentially useful for performing multiple ECG processing tasks using one set of preprocessing filters.
Publication Types:
PMID: 9932341 [PubMed - indexed for MEDLINE]