Your browser version may not work well with NCBI's Web applications. More information here...
Related Articles, Links
Click here to read
A comparison of the noise sensitivity of nine QRS detection algorithms.

Friesen GM, Jannett TC, Jadallah MA, Yates SL, Quint SR, Nagle HT.

Auburn University, AL 36830.

The noise sensitivities for nine different QRS detection algorithms were measured for a normal, single-channel lead II, synthesized ECG corrupted with five different types of synthesized noise. The noise types were electromyographic interference, 60 Hz powerline interference, baseline drift due to respiration, abrupt baseline shift, and a composite noise constructed from all of the other noise types. The percentage of QRS complexes detected, the number of false positives, and the detection delay were measured. None of the algorithms were able to detect all QRS complexes without any false positives for all of the noise types at the highest noise level. Algorithms based on amplitude and slope had the highest performance for EMG-corrupted ECG. An algorithm using a digital filter had the best performance for the composite noise corrupted data.

Publication Types:
PMID: 2303275 [PubMed - indexed for MEDLINE]