Table 2

Parameter variation test results

Parameters
Perfect valve
Diameter test
Stiffness test
Distribution test
Length test
Location test

Diameter (cm)
1.2
0.8
1.6
1.2
1.2
1.2
1.2
Length (m)
1
1
1
1
0.50
0.75
1
Young's modulus (kPa)
1000
1000
500
2000
1000
1000
1000
Poisson's ratio
0.5
0.5
0.5
0.5
0.5
0.5
Blood viscosity (mPas)
4
4
4
4
4
4
Blood density (Kgm-3)
1000
1000
1000
1000
1000
1000
Gravity application time (s)
Near instantaneous
Near instantaneous
Near instantaneous
Near instantaneous
Near instantaneous
Near instantaneous
Valve distribution
One
One
One
Two
One
One
Valve location from inlet (m)
0.5
0.5
0.5
0.25 and 0.75
0.5
0.25
0.75
Valve performance
Perfect
Perfect
Perfect
Perfect
Perfect
Perfect
First peak (mmHg)
60.2 (8.01 kPa)
66.3 (8.81 kPa)
56.8 (7.55 kPa)
56.3 (7.48 kPa)
63.2 (8.4 kPa)
32.6 (4.34 kPa)
33.2 (4.41 kPa)
46.8 (6.23 kPa)
32.6 (4.33 kPa)
87.2 (11.6 kPa)
Second peak (mmHg)
81.9 (10.9 kPa)
114.3 (15.2 kPa)
74.9 (9.97 kPa)
74.6 (9.92 kPa)
95.5 (12.7 kPa)
55.6 (7.39 kPa)
55.9 (7.44 kPa)
57 (7.58 kPa)
99.3 (13.2 kPa)
90.2 (12 kPa)
Maximum pressure (mmHg)
93.3 (12.4 kPa)
115.8 (15.4 kPa)
89.5 (11.9 kPa)
90.9 (12.1 kPa)
104.5 (13.9 kPa)
92.9 (12.36 kPa)
57.2 (7.61 kPa)
77.5 (10.31 kPa)
110.5 (14.7 kPa)
105.3 (14 kPa)
Dynamic pressure ratio
1.25
1.47
1.22
1.21
1.41
1.25
1.53
1.39
1.49
1.41
Valve closed time (s)
0.17
0.11
0.19
0.26
0.11
0.09 and 0.17
0.06
0.11
0.16
0.15
Maximum collapse (%)
20.2
3.64
24.5
48.2
6.06
35.6
3.58
10.6
3.25
42.3
Maximum expansion (%)
9.89
12.1
9.87
21.2
5.41
9.89
5.92
7.19
11.9
11.5

Zervides et al. BioMedical Engineering OnLine 2008 7:8   doi:10.1186/1475-925X-7-8