Table 1

Endurance testing of the Biomet Matthews Nail and the Dynamic Compression Screw tested with Y fractured composite femurs


Biomet Matthews Nail with Y fractured femur
Dynamic Compression Screw with Y fractured femur

Max. & Min. of Load Cycle (kN)
Specimen Number
Cycles Completed
Specimen Number
Cycles Completed

2.0 & 0.2
1
1,000,000
1
1,000,000#
2.0 & 0.2
2
315,043
2
1,000,000#
3.0 & 0.3
3
22,356+


3.0 & 0.3
4
78,386



# Post test examination revealed a fracture through the second hole, superior to the lag screw

+Lateral condyle bone segment shifted laterally until purchase on the distal locking screw became sufficient to bend it

O'Connor-Read et al. BioMedical Engineering OnLine 2008 7:3   doi:10.1186/1475-925X-7-3