List of timers used in the AF-VP model |
||
| Timer |
Start condition |
Stop condition |
|
|
||
| AF interval timer |
Arrival of AF impulse |
End of AF interval |
| RR interval timer |
Any VS or VP event |
Next VS or VP event |
| AVJ refractory timer |
AVJ activation |
End of AVJ refractory period |
| VP clock timer |
Any VS or VP event |
End of VP interval |
| Antegrade AVJ timer |
Antegrade AVJ activation |
End of antegrade AVD |
| Retrograde AVJ timer |
Retrograde AVJ activation |
End of retrograde AVD |
| Antegrade ventricle timer |
End of antegrade AVD |
VS event or ventricle fusion |
| Retrograde ventricle timer |
Delivery of VP |
Retrograde invasion of AVJ or ventricle fusion |
Lian et al. BioMedical Engineering OnLine 2007 6:9 doi:10.1186/1475-925X-6-9 |
||