Statistical overview of rate of success of the real-time neuro-fuzzy system for sEMG pattern discrimination system |
|||||||
| Movements |
|||||||
|
|
|||||||
| Participated subjects |
Opening |
Closing |
Wrist flexion |
Wrist extension |
Pinch |
Thumb flexion |
Average Results for 6 movements |
|
|
|||||||
| Subject_1 |
98% |
100% |
94% |
94% |
96% |
92% |
95.67% |
| Subject_2 |
98% |
100% |
100% |
96% |
98% |
94% |
97.67% |
| Subject_3 |
96% |
100% |
88% |
98% |
98% |
92% |
95.33% |
| Subject_4 |
100% |
100% |
94% |
96% |
100% |
98% |
98% |
| Average results |
98% |
100% |
94% |
96% |
98% |
94% |
96.67% |
| Standard deviation (STD) for each movements |
1.41% |
0% |
4.2% |
4.2% |
1.41% |
2.45% |
1.18% |
Khezri and Jahed BioMedical Engineering OnLine 2007 6:45 doi:10.1186/1475-925X-6-45 |
|||||||