BioMedical Engineering OnLine
|
Viewing options:Associated material:Related literature:- Articles citing this article
- Other articles by authors
- Related articles/pages
Tools:Post to:
|
 ResearchFinite Element Analysis of Hepatic Radiofrequency Ablation Probes using Temperature-Dependent Electrical ConductivityIsaac Chang  Office of Science and Technology, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Rockville MD USA author email corresponding author email
BioMedical Engineering OnLine 2003,
2:12doi:10.1186/1475-925X-2-12 Abstract
Background
Few finite element models (FEM) have been developed to describe the electric field, specific absorption rate (SAR), and the temperature distribution surrounding hepatic radiofrequency ablation probes. To date, a coupled finite element model that accounts for the temperature-dependent electrical conductivity changes has not been developed for ablation type devices. While it is widely acknowledged that accounting for temperature dependent phenomena may affect the outcome of these models, the effect has not been assessed.
Methods
The results of four finite element models are compared: constant electrical conductivity without tissue perfusion, temperature-dependent conductivity without tissue perfusion, constant electrical conductivity with tissue perfusion, and temperature-dependent conductivity with tissue perfusion.
Results
The data demonstrate that significant errors are generated when constant electrical conductivity is assumed in coupled electrical-heat transfer problems that operate at high temperatures. These errors appear to be closely related to the temperature at which the ablation device operates and not to the amount of power applied by the device or the state of tissue perfusion.
Conclusion
Accounting for temperature-dependent phenomena may be critically important in the safe operation of radiofrequency ablation device that operate near 100°C. |