Log on/register
BioMed Central home | Journals A-Z | Feedback | Support | My details
 
Open AccessHighly AccessResearch

Finite Element Analysis of Hepatic Radiofrequency Ablation Probes using Temperature-Dependent Electrical Conductivity

Isaac Chang email

Office of Science and Technology, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Rockville MD USA

author email corresponding author email

BioMedical Engineering OnLine 2003, 2:12doi:10.1186/1475-925X-2-12

Published: 8 May 2003

Abstract

Background

Few finite element models (FEM) have been developed to describe the electric field, specific absorption rate (SAR), and the temperature distribution surrounding hepatic radiofrequency ablation probes. To date, a coupled finite element model that accounts for the temperature-dependent electrical conductivity changes has not been developed for ablation type devices. While it is widely acknowledged that accounting for temperature dependent phenomena may affect the outcome of these models, the effect has not been assessed.

Methods

The results of four finite element models are compared: constant electrical conductivity without tissue perfusion, temperature-dependent conductivity without tissue perfusion, constant electrical conductivity with tissue perfusion, and temperature-dependent conductivity with tissue perfusion.

Results

The data demonstrate that significant errors are generated when constant electrical conductivity is assumed in coupled electrical-heat transfer problems that operate at high temperatures. These errors appear to be closely related to the temperature at which the ablation device operates and not to the amount of power applied by the device or the state of tissue perfusion.

Conclusion

Accounting for temperature-dependent phenomena may be critically important in the safe operation of radiofrequency ablation device that operate near 100°C.


© 1999-2010 BioMed Central Ltd unless otherwise stated. Part of Springer Science+Business Media.