sEMG feature evaluation for identification of elbow angle resolution in graded arm movement

Abstract

Automatic and accurate identification of elbow angle from surface electromyogram (sEMG) is essential for myoelectric controlled upper limb exoskeleton systems. This requires appropriate selection of sEMG features, and identifying the limitations of such a system.

This study has demonstrated that it is possible to identify three discrete positions of the elbow; full extension, right angle, and mid-way point, with window size of only 200 milliseconds. It was seen that while most features were suitable for this purpose, Power Spectral Density Averages (PSD-Av) performed best. The system correctly classified the sEMG against the elbow angle for 100% cases when only two discrete positions (full extension and elbow at right angle) were considered, while correct classification was 89% when there were three discrete positions. However, sEMG was unable to accurately determine the elbow position when five discrete angles were considered. It was also observed that there was no difference for extension or flexion phases.

Background

Exoskeleton systems of the arm have number of applications such as support for the elderly, defense personnel, and people with skeletal injuries [13]. For effective application of these devices, the user should be able to control them naturally and intuitively. While there are number of options for commanding such systems such as the use of mechanical sensors, brain computer interface and use of surface electromyogram (sEMG) of the associated muscles, sEMG provides a natural and intuitive interface for the user [39]. This can also offer the user with proportional control where the exoskeleton device can follow the body movement. However, the difficulty with such sEMG based controlling strategies is the poor sensitivity and specificity, leading to poor reliability.

The angle of the elbow is an important command of the exoskeleton of the upper limb. To obtain this from sEMG recording requires the appropriate selection of sEMG features which then have to be classified to identify the elbow angle. Different researchers have used different features [1014]. However, none of the researchers have performed a comparative between all of these features.

Proportional control requires the system to identify the position of the body based on the sEMG of the effective muscles. While some simple systems provide binary resolutions such as flexion and extension commands, this does not offer proportional control, and is not intuitive [1518]. There is the need for higher resolution where the user is able to give finer commands to the exoskeleton device for performing the upper limb actions. Higher resolution requires the classification system to have larger number of classes. However, there is a tradeoff between the number of classes and the system accuracy and reliability, and there is the need for determining the relationship between the number of classes and the system accuracy.

The aim of this research was to determine the relationship between the classification system sensitivity, specificity and accuracy for different resolutions of the elbow angles (number of classes or number of arm positions), and determine the feasibility of high resolution identification of the elbow angle. A comparison was performed between the commonly used features of sEMG to select the most suitable feature set for the proposed classification system. The accuracy, sensitivity and specificity of each of these features in the recognition of the elbow angle were obtained. The relationship between resolution (number of classes) and the accuracy, sensitivity and specificity of recognition of the position of the arm was studied.

Methods

The experimental protocol was approved by the Research Ethics Committee from São Judas University, São Paulo, Brazil, by letter; COEP-USJT-No.076-2010, and in accordance with the Helsinki accord (modified 2004). Seven able-bodied volunteer subjects (4 men and 3 women), average age 34.6, participated in the experiment. They were verbally and in writing explained the purpose of the experiments and the experimental protocol, and experiments were performed after obtaining their verbal and written informed consent. Before recording the data, the participants were allowed sufficient time to familiarize themselves with the equipment and the protocol. Multiple trial runs were performed till the volunteers were comfortable with the experiment.

Experiments

Equipment

A custom-made elbow angle monitoring device (Figure 1) was used to monitor elbow angular position. This device restricts the movement of the arm at the elbow in the horizontal plane and is fixed at the height of shoulder of the subject. A goniometer records the angle between the upper arm and the forearm at the elbow. The users were given visual feedback of the elbow angle on the screen throughout the experiment.

Figure 1
figure 1

Custom made elbow angle monitoring device.

Two channel sEMG signals were recorded using Powerlab (AdInstruments), using disposable pre-gelled bipolar surface electrodes (Noraxon). The ground electrode was placed on the acromion point. The electrodes, for recording the biceps sEMG, were placed above the motor point of the short head, on the line between the acromion and the fossa cubit, at 1/3 from the fossa cubit. And the electrodes, for the lateral head of the triceps, were placed on the middle of the line between the posterior crista of the acromion and the olecranon at two finger widths lateral to the line. The inter-electrode was maintained at 20 mm (center to center). The signal was sampled at 1000 Hz/channel and filtered using eighth-order, switched-capacitance, Bessel type filter, range 20–500 Hz and notch at 60 Hz.

Experimental protocol

During the experiments, the angle of the elbow along with the sEMG from the biceps brachii and triceps brachii was recorded. The participants were given continuous visual feedback of the angle of the elbow.

During the experiment, the participant performed graded flexion/extension movements with 10° shifts every 3 s, going from full extension position (0°) to 90° of flexion and returning to the full extension position in the same way (Figure 2). The participants were provided with audio cues for timing the movement. This procedure was repeated 3 times for each volunteer, with 5 minutes rest period between experiments to ensure there was no fatigue.

Figure 2
figure 2

Graded movement for arm flexion/ extension showing elbow angular position as a function of time.

Signal processing

There are two possible techniques for determining the position of the arm from sEMG recordings of the biceps and triceps muscles; dynamic or static. The dynamic relates to the sEMG recorded when the arm is in motion and the muscle is producing the motion, while the static is when the arm is not moving and the muscle activity is isometric. In this situation, contractions above certain threshold are usually used, being stronger than those used in dynamic movement without load [5, 8, 9].

Segmentation

Good classification of the signal requires high signal to noise ratio. While isometric contraction during relaxed, maintained position of the arm has very low muscle activity because of which the signal to noise ratio becomes very poor, sEMG during the movement is significantly higher, with higher signal to noise ratio. Thus, dynamic contraction phase during arm movement was selected for the purpose of signal analysis.

Researchers have identified delays need to be less than 250 ms for user satisfaction. Analysis of the signal also showed that first 200 ms of each step movement [10] is consistent and hence was considered for analysis. The signal was segmented; the 200 ms at the start of each 10° shift movement was selected forming one of the data vectors in each data class, and is indicated by two examples shown by the red regions in Figure 2. The signal was normalized based on the maximum value in this range. It was then labeled based on the angle of the elbow such as Bf10 being the 200 ms segment of sEMG of the biceps obtained at the completion of the 10° flexion. Experimental class set ups were defined in Table 1.

Table 1 Angular interval variation for each class set up, for each movement phase

Feature extraction

Appropriate section of features to represent the sEMG signal is essential for accurate identification of actions and movements [11, 19]. While researchers have tested the efficacy of number of features, few publications have compared the accuracy, sensitivity and specificity of different features. In this work, 9 commonly reported features were extracted. These are briefly described in Table 2, where, x k is the k th sample of a total of N, in the window i of a total of I number of windows [10, 1214].

Table 2 Feature definition

Linear discriminant analysis as classifier

Linear Discriminant Analysis (LDA) is a statistical method based on linear transformations of the data set, projected onto the directions that achieve the best class separability. The goal is to maximize the between-class scatter matrix while minimizing the within-class scatter matrix [20, 21].

W l d a =arg max W W T S b W W T S w W
(1)

According to Fisher criterion, the solution for the Equation 1, that defines the projection matrix W lda , can be achieved as a typical problem of eigenvectors, which the solution are the eigenvectors and the eigenvalues of S b S w 1 , with at most (g − 1) nonzero eigenvalues, where g is the number of classes [20, 21].

However, in practical applications, the within-class scatter matrix S w can be singular. This comes from the fact that, in general, the number of patterns in the training set N i is much smaller than the dimensionality d of the data set [20, 21]. To deal with this singularity problem, one of the methods in the literature, known as Regularized LDA (RLDA), adds a constant α to the diagonal elements of the pooled matrix S p (defined by Equation 2), where α is known as the regularization parameter. In this work, α ranged from 10-9 to 0.9, interval defined in a previous study for class separability purposes [22].

S P = S w N g
(2)

The analysis was first done when all the subjects were pooled together during training and testing. However, the results from the leave one out technique were very poor, and this approach was discarded. Subsequently, the training was repeated for each subject individually, and these results have been shown in this paper. This also demonstrates that there is significant variation between subjects, and indicates that it is important for the classifier to be trained for each user.

Each feature (Table 2) applied over the 200 ms segment produced a vector corresponding to each of the two muscles; biceps and triceps. These feature vectors were the input to the LDA. Leave One Out method was used to validate the system, where the data set was divided in groups of 2 samples for each class and each subject for training the LDA, while the remaining sample was used for testing for each subject. This was repeated three times to ensure there was no bias. The average of the three trials was obtained and is indicative of the ability for the system to determine the elbow angle, taking into account the differences between multiple samples and the considered class set ups.

Results

The average accuracy, sensitivity and specificity achieved by each feature, for each movement phase, and for each resolution (class set up) are shown in Table 3. From this table it is observed that for resolution = 2 classes, PSD has the highest accuracy, sensitivity and specificity for both movement directions (flexion and extension), while ZC has the lowest.

Table 3 Performance metrics (Sesensitivity%; Spspecificity%; Accaccuracy%)

From this table it is also observed that PSD-Av also has high sensitivity, specificity and accuracy for classification of 2 (full extension and 90° flexion) and 3-class (including a half-way position class) set ups. However, for the configuration using 3 classes, during extension phase, the accuracy is 84%, sensitivity = 84%, and specificity = 92%. Such a system may not be suitable for applications where any error in recognizing the command may cause injury to the user.

The results also show that the response of the system to classify 5-class set up decreases, which shows that no sEMG feature of the biceps and triceps is suitable for accurately identifying the elbow angle for exoskeleton control for higher resolutions. The results were also confirmed by the Kappa based comparative statistics [23] to show the interobserver variation and reported in Table 4. Based on the Cohen’s Kappa statistic value from Table 4 and the study by Viera and Garret [23] it is shown that the Kappa value above 0.60 suggests the substantial agreement with the predicted and actual observation.The scatter plots of the normalized PSD-Av of the triceps vs biceps are shown in Figure 3. PSD-Av is a feature which consists of 16 parameters for each muscle, and was selected because statistical analysis confirmed it to have the lowest error.Figure 3(a) is a plot for 3 classes; 0°-10°, 40°-50° and 80°-90° while, Figure 3(b) is a plot for 5 classes; 0°-10°, 20°-30°, 40°-50°, 60°-70° and 80°-90°. From these plots, it is observed that there are significant differences in the 0°-10°, 40° -50° and 80°-90° classes, but there is significant overlap when two additional classes; 20°-30°, and 60°-70° are added. This demonstrates the limitation of such an approach for myoelectric exoskeleton systems.

Table 4 Kappa statistic results
Figure 3
figure 3

Example of scatter- plots of normalized PSD- Av data values during the flexion movement phase for: (a) 3- class set up and (b) 5- class set up.

Discussion and conclusions

The results show that sEMG system can be used effectively to identify the flexion and extension of the elbow when we consider two state situations; arm at full extension, and arm at 90° flexion. This work has also shown that when the number of classes of classification increased to 3 classes, the system accuracy dropped, with sensitivity = 84% and specificity = 92%. Such a system may be suitable for limited applications due to the relatively low sensitivity, which could cause injury to the user.

When the number of classes was increased to 5-class set up, the error was higher compared with the situation of 2 and 3-class set ups, with sensitivity = 64%, and specificity = 91% in the best case. This indicates that sEMG classification is suitable for the identification of small number of elbow positions but unsuitable for being used for high resolution conditions. Poor sensitivity will frustrate the user and make the system non-functional. One reason for poor sensitivity may be due to the narrow window of 200 ms. However, this is essential because earlier studies have identified that delays greater than 250 ms causes observable delays to the user, and can be the cause of errors. The other reason is the similarities between consecutives positions due to its discretization.

The results also showed that at small number of elbow positions, the performance of most features was similar, but as the number become higher, none of them achieved reasonable results. While relating sEMG with angles in between the extreme flexion and extension did not give good results, however, it should be noted that nil error in detecting full flexion and extension levels is not comparable with relatively higher errors during the in-between steps. But the relatively large error highlights the relatively limited applications of such an approach.

There are other options that may be considered as an alternate to a classification of small number of sEMG channels; high density myoelectric recordings, mechanical sensor system, a hybrid system, or the use of an intelligent myoelectric system where model based approach may be used. While mechanical sensor systems have number of shortcomings, the hybrid system that combines the sEMG with the mechanical sensors may reduce the errors while providing the user with natural and intuitive control. Another approach, where the mechanical sensing may not be possible, is to develop an intelligent system modeling the movement, such as is trained to estimate the speed of the action of the user. In this approach, the system would be trained to estimate the time of the action for an individual and assuming the movement to be continuous. Such a system could provide an alternate for a myoelectric based proportional controller system also known as tracking systems.

References

  1. Herr H: Exoeskeleton and orthoses classification, design challenges and future directions. J Neuro Eng Rehabil 2009, 6: 21. doi:10.1186/1743–0003–6-21 10.1186/1743-0003-6-21

    Article  Google Scholar 

  2. Lo HS, Xie SQ: Exoskeketon robots for upper limb rehabilitation: state of the art and future prospects. Med Eng Phys 2012, 34: 261–268. 10.1016/j.medengphy.2011.10.004

    Article  Google Scholar 

  3. Kiguchi K, Fukuda T: Upper-Limb Exoskeletons for Physically Weak Persons. In Rehabilitation Robotics, August, Chap. 16. Edited by: Kommu SS. Viena, Austria: InTech; 2007:287–299. . http://www.intechopen.com/books/rehabilitation_robotics/upper-limb_exoskeletons_for_physically_weak_persons

    Google Scholar 

  4. Li G: Electromyography Pattern Recognition Based Control of Powered Multifunctional Upper Limb Prostheses. In Advances in Applied Electromyography, Chap. 6. Edited by: Mizrahi J. Rijeka, Croatia: InTech; 2011:99–117. http://www.intechopen.com/books/advances-in-applied-electromyography/electromyography-pattern-recognition-based-control-of-powered-multifunctional-upper-limb-prostheses

    Google Scholar 

  5. Scheme E, Englehart K: Electromyogram pattern recognition for control of powered upper-limb prostheses: State of the art and challenges for clinical use. J Rehabil Res Dev 2011,48(6):643–660. 10.1682/JRRD.2010.09.0177

    Article  Google Scholar 

  6. Lenzi T, De Rossi S, Vitiello N, Carrozza M: Intention-based EMG control for powered exoskeletons. IEEE Trans Biomed Eng 2012,59(8):2180–2190.

    Article  Google Scholar 

  7. Kiguchi K, Hayashi Y: An EMG-based control for an upper-limb power-assist exoskeleton robot. IEEE Trans Syst Man Cybern B Cybern 2012,42(4):1064–1071.

    Article  Google Scholar 

  8. Zhang X, Zhou P: High-density myoelectric pattern recognition toward improved stroke rehabilitation. IEEE Trans Biomed Eng 2012,59(6):1649–1657.

    Article  Google Scholar 

  9. Muceli S, Farina D: Simultaneous and proportional estimation of hand kinematics from emg during mirrored movements at multiple degrees-of-freedom. IEEE Trans Neural Syst Rehabil Eng 2012,20(3):371–378.

    Article  Google Scholar 

  10. Hudgins B, Parker P, Scott RN: A new strategy for multifunction myoelectric control. IEEE Trans Biomed Eng 1993,40(1):82–94. 10.1109/10.204774

    Article  Google Scholar 

  11. Englehart K, Hudgins B, Parker PA, Stevenson M: Classification of the myoelectric signal using time-frequency based representations. Med Eng Phys 1999,21(6–7):431–438.

    Article  Google Scholar 

  12. Liu X, Zhou R, Yang L, Li G: Performance of Various EMG Features in Identifying arm Movements for Control of Multifunctional Prostheses. In Proceedings of the IEEE Youth Conference on Information Computing and Telecommunication. Beijing: IEEE; 2009:287–290.

    Google Scholar 

  13. Oskoei MA, Hu H: GA-Based Feature Subset Selection for Myoelectric Classification. In Proceedings of the IEEE International Conference on Robotics and Biomimetics. Kunming, China: IEEE; 2006:1465–1470.

    Google Scholar 

  14. Du S, Vuskovic M: Temporal vs. Spectral Approach to Feature Extraction from Prehensile EMG Signals. In Proceedings of the IEEE International Conference on Information Reuse and Integration-IRI. Las Vegas, USA: IEEE; 2004:344–350.

    Google Scholar 

  15. Artemiadis PK, Kyriakopoulos KJ: An EMG-based robot control scheme robust to time-varying EMG signal features. IEEE Trans Inf Technol Biomed 2010,14(3):582–588.

    Article  Google Scholar 

  16. Alkan A, Gunay M: Identification of EMG signals using discriminant analysis and SVM classifier. Expert Syst Appl 2012,39(1):44–47. 10.1016/j.eswa.2011.06.043

    Article  Google Scholar 

  17. Farfán FD, Politti JC, Felice CJ: Evaluation of EMG processing techniques using information theory. BioMedical Eng Online 2010, 9: 72. http://www.biomedical-engineering-online.com/content/9/1/72 10.1186/1475-925X-9-72

    Article  Google Scholar 

  18. Khokhar ZO, Xiao ZG, Menon C: Surface EMG pattern recognition for real-time control of a wrist exoskeleton. BioMedical Eng Online 2010, 9: 41. http://www.biomedical-engineering-online.com/content/9/1/41 10.1186/1475-925X-9-41

    Article  Google Scholar 

  19. Bunderson NE, Kuiken TA: Quantification of feature space changes with experience during electromyogram pattern recognition control. IEEE Trans Neural Syst Rehabil Eng 2012,20(3):239–246.

    Article  Google Scholar 

  20. Belhumeur P, Hespanha P, Kriegman D: Eigenfaces vs fisherfaces: recognition using class specific linear projection. IEEE Trans Pattern Anal Mach Intell 1997,19(7):711–720. 10.1109/34.598228

    Article  Google Scholar 

  21. Thomaz CE, Gillies DF: A Maximum Uncertainty LDA-Based Approach for Limited Sample Size Problems With Application to Face Recognition. In Proceedings of the XVIII Brazilian Symposium on Computer Graphics and Image Processing-SIBGRAPI’05. Natal, Brazil: IEEE; 2005:89–96.

    Chapter  Google Scholar 

  22. Castro MCF: Statistical Approach for Angular Position Separability Classes of EMG Data. In Proceedings of the ISSNIP Biosignals and Biorobotics Conference-BRC. Vitória, Brazil: IEEE; 2011:1–6. doi:10.1109/BRC.2011.5740663

    Chapter  Google Scholar 

  23. Garrett JA, Viera JM: Understanding interobserver agreement: the kappa statistic. Fam Med 2005,37(5):360–363.

    Google Scholar 

Download references

Acknowledgements

The authors would like to thank Centro Universitário da FEI, FAPESP and RMIT University for their support in this project.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Maria Claudia F Castro.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

MCFC planned and conducted the experiments. ELC performed signal analysis. PTAJ wrote the draft of the paper. SPA performed data analysis. DKK wrote the paper. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Castro, M.C.F., Colombini, E.L., Junior, P.T.A. et al. sEMG feature evaluation for identification of elbow angle resolution in graded arm movement. BioMed Eng OnLine 13, 155 (2014). https://doi.org/10.1186/1475-925X-13-155

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1475-925X-13-155

Keywords